This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

HEPA FILTER 100105-01 (2016-04-29)

Starting date:
April 29, 2016
Posting date:
May 31, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-58628

Affected products

A. HEPA FILTER 100105-01

Reason

Filter contained a two part urethane sealant that was not mixed with the proper ratio to ensure a full cure. It was discovered that the uncured sealant can leak out and drip inside the drier at higher temperatures.

Affected products

A. HEPA FILTER 100105-01

Lot or serial number

N/A

Model or catalog number

N/A

Companies
Manufacturer

Natus Medical Incorporated,

5900 First Avenue South,

Seattle

98108

Washington

UNITED STATES