HEMOSPRAY ENDOSCOPIC HEMOSTAT (2019-07-24)
- Starting date:
- July 24, 2019
- Posting date:
- July 30, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type I
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-70589
Last updated: 2019-07-30
Affected Products
HEMOSPRAY ENDOSCOPIC HEMOSTAT
Reason
This product is being recalled because Cook has received eight (8) complaints of the Hemospray device being unable to spray powder due to misassembly of devices. This has been reported to have led to an inability to achieve hemostasis, and in one case, the patient was transferred to surgery. There is also a potential risk of death if hemostasis is unable to be achieved in emergent cases. As stated above, the devices may not spray powder resulting in an inability to achieve hemostasis
Affected products
HEMOSPRAY ENDOSCOPIC HEMOSTAT
Lot or serial number
W4179646
W4180858
W4180861
W4181073
Model or catalog number
HEMO-10
Companies
- Manufacturer
-
Wilson-Cook Medical Inc.
4900 Bethania Station Rd.
Winston Salem
North Carolina
UNITED STATES