Health product recall

HEMOSPRAY ENDOSCOPIC HEMOSTAT (2019-07-24)

Starting date:
July 24, 2019
Posting date:
July 30, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-70589


Last updated:

Affected Products

HEMOSPRAY ENDOSCOPIC HEMOSTAT

Reason

This product is being recalled because Cook has received eight (8) complaints of the Hemospray device being unable to spray powder due to misassembly of devices. This has been reported to have led to an inability to achieve hemostasis, and in one case, the patient was transferred to surgery. There is also a potential risk of death if hemostasis is unable to be achieved in emergent cases. As stated above, the devices may not spray powder resulting in an inability to achieve hemostasis

Affected products

HEMOSPRAY ENDOSCOPIC HEMOSTAT

Lot or serial number

W4179646

W4180858

W4180861

W4181073

Model or catalog number

HEMO-10

Companies
Manufacturer

Wilson-Cook Medical Inc.

4900 Bethania Station Rd.

Winston Salem

North Carolina

UNITED STATES