Health product recall

Heater-Cooler Unit HCU 40 (2018-08-03)

Starting date:
August 3, 2018
Posting date:
August 17, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67454

Affected products

Heater-Cooler Unit HCU 40

Reason

The vacuum valve is located in both, the patient and the cardioplegia water circuit. Maquet cardiopulmonary has received complaints in relation to device malfunctions caused by leaking vacuum valves. This issue is serial number specific and only 3 affected serial numbers were sold in Canada. The bacteriostatic agent used in the storage and distribution phase of manufacturing compromised the tightness of the vacuum valve.  This has been identified and changed.

Affected products

Heater-Cooler Unit HCU 40

Lot or serial number
  • S/N 90440575
  • S/N 90440576
  • S/N 90440577
Model or catalog number

70104.4054 (200-240V)

Companies
Manufacturer
Maquet Cardiopulmonary GmbH
Kehler Str. 31
Rastatt
76437
GERMANY