This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

HEATER COOLER UNIT HCU 40 ( 2015-07-02)

Starting date:
July 2, 2015
Posting date:
July 21, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-54306

Affected Products

A.HEATER COOLER UNIT HCU 40

Reason

Maquet Cardiopulmonary recently detected a leak in the water circuit of an HCU 40 during the final quality inspection. The investigation revealed that the leakage was caused by a damaged sealant in the water circuit. The damage occurred during device assembly and it cannot be ruled out with absolute certainty that HCU 40s in the field equipped with the same type of sealant could exhibit this failure as well. The HCU 40 with the serial number 90440265 could potentially exhibit a leak in the water circuit. In case of water leakage a short circuit of an electrical component and device shut down are the potential consequences.

Affected products

A.HEATER COOLER UNIT HCU 40

 

Lot or serial number

90440265

Model or catalog number

70105.4054 (200-240V)

Companies
Manufacturer

Maquet Cardiopulmonary Ag

Kehler Str. 31

Rastatt