Health product recall

HEATER-COOLER UNIT 30 SYSTEM - MAIN UNIT (2015-10-28)

Starting date:
October 28, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-55758

Affected Products

HEATER-COOLER UNIT 30 SYSTEM - MAIN UNIT

Reason

MAQUET Cardiopulmonary has received complaints associated with a failure of the power supply board of the HCU 30. An internal investigation showed that a drop of more than 10% voltage can cause an overheating and subsequent failing of the NTC (negative temperature coefficient) R56 thermistor of the power supply board. This failure typically occurs when the compressor is switched on, i.e. before patient connection. A defective thermistor results in an inoperable compressor or in an inoperable HCU 30. There are no complaints reported associated with a patient injury or death.

Affected products

HEATER-COOLER UNIT 30 SYSTEM - MAIN UNIT

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 00705501
  • 00939001
  • 0704629
Companies
Manufacturer

MAQUET CARDIOPULMONARY GMBH

KEHLER STR. 31, RASTATT, GERMANY, 76437

RASTATT

76437

GERMANY