This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

HEATER COOLER SYSTEM 3T (2016-06-21)

Starting date:
June 21, 2016
Posting date:
July 19, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Hospitals, Healthcare Professionals
Identification number:
RA-59390

Affected products

A. HEATER COOLER SYSTEM 3T

Reason

LivaNova has determined that certain oxygenators containing a heat-exchanger with a polymer membrane separating the blood compartment from the water compartment, when used with the cleaning an disinfection regimen described in the IFU for the heater cooler system 3T (IFU versions applicable since 2013), may result in diffusion of a small quantity of hydrogen peroxide from the water side to the blood side during a typical procedure, in a quantity that exceeds the allowable limits determined by LivaNova.

Affected products

A. HEATER COOLER SYSTEM 3T

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number
  • "16-02-81
  • 16-02-82
  • 16-02-85"
Companies
Manufacturer

Sorin Group Deutschland GMBH

LINBERGHSTRASSE 25

MUNICH

80939

GERMANY