HEARTWARE VENTRICULAR ASSIST SYSTEM (2021-03-04)
- Starting date:
- March 4, 2021
- Posting date:
- March 25, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-75201
Last updated:
2021-03-25
Affected Products
- HEARTWARE VENTRICULAR ASSIST SYSTEM - HEARTWARE CONTROLLER
- HEARTWARE VENTRICULAR ASSIST SYSTEM - HEARTWARE CONTROLLER AC ADAPTER
- HEARTWARE VENTRICULAR ASSIST SYSTEM - HEARTWARE CONTROLLER DC ADAPTER CONTROLLER
- HEARTWARE VENTRICULAR ASSIST SYSTEM - ALARM ADAPTER
- HEARTWARE VENTRICULAR ASSIST SYSTEM - DATA CABLE
- HEARTWARE VENTRICULAR ASSIST SYSTEM - HEARTWARE BATTERY
Reason
Medtronic is informing customers of an issue related to the Heartware HVAD Controller. Specifically, there is the potential for the Controller Power Cables (AC adapter, DC adapter, and battery cables), HVAD alarm adapter, and HVAD monitor data cables to cause damage to the HVAD Controller Ports when inserting the Cables into the Controller Ports, due to misalignment of the Cables to the port. Over time, repeated misalignment causes wear on the power Cables and Data Cables and may cause damage or bending on the Controller Port metal pins. This damage can result in the inability to fully lock the Cable Connectors into the Controller, potentially causing interruptions or disconnections in power to the Controller, which may result in Controller Power loss and HVAD pump stop or loss of communication to the HVAD monitor. When this occurs, it may be necessary to exchange the Controller to restore connections between the power sources and Controller.
To reduce the risk of damage to the HVAD Controller Power Cables and Data Cables that can be caused by the misalignment, Medtronic is providing more detailed instructions in an updated Instruction For Use (IFU) and Patient Manual (PM) about the criteria and frequency for inspecting the Cables and Controller Ports.
Affected products
A. HEARTWARE VENTRICULAR ASSIST SYSTEM - HEARTWARE CONTROLLER
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
1403US
1407CA
1420
Companies
- Manufacturer
-
Heartware Inc.
14400 NW 60th Avenue
Miami Lakes
33014
Florida
UNITED STATES
B. HEARTWARE VENTRICULAR ASSIST SYSTEM - HEARTWARE CONTROLLER AC ADAPTER
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
1430CA
C. HEARTWARE VENTRICULAR ASSIST SYSTEM - HEARTWARE CONTROLLER DC ADAPTER CONTROLLER
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
1440
D. HEARTWARE VENTRICULAR ASSIST SYSTEM - ALARM ADAPTER
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
1450
E. HEARTWARE VENTRICULAR ASSIST SYSTEM - DATA CABLE
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
1575
F. HEARTWARE VENTRICULAR ASSIST SYSTEM - HEARTWARE BATTERY
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
1650DE