Health product recall

Heartstart XL, Heartstart MRX Defibrillator/Monitor and Heartstart XL+ (2018-04-18)

Starting date:
April 18, 2018
Posting date:
May 18, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66814

Affected products

A. Heartstart XL
B. HEARTSTART MRX Defibrillator/Monitor
C. HEARTSTART XL+

Reason

When a Philips monitor/defibrillator is receiving an ECG signal from an auxiliary bedside monitor via a sync cable or ECG out cable, the following can occur if the monitor/defibrillator experiences interference from
electrical fast transients (EFTs):
- On the HeartStart MRx and HeartStart XL, EFT noise may be misinterpreted as an R-wave.
- On the HeartStart XL+, EFT noise can disable ECG monitoring and potentially interrupt demand mode pacing*.
*Note: It is contrary to the XL+ Instructions for Use to perform demand mode pacing while using the ECG out cable or obtaining the ECG signal from a bedside monitor. The XL+ Instructions for Use includes the following warning: "When pacing in Demand Mode, the ECG cable from the patient must be directly connected to the HeartStart XL+." If the user follows this warning, this problem cannot occur on the XL+.

Affected products

A. Heartstart XL

Lot or serial number

Not applicable

Model or catalog number

M4735A

Companies
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
UNITED STATES

B. HEARTSTART MRX Defibrillator/Monitor

Lot or serial number

Not applicable

Model or catalog number

M3535A
M3536A

Companies
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
UNITED STATES

C. HEARTSTART XL+

Lot or serial number

Not applicable

Model or catalog number

861290

Companies
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
UNITED STATES