HeartStart MRX Rechargeable Lithium Ion Battery (2018-10-27)
- Starting date:
- October 27, 2018
- Posting date:
- November 30, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-68466
Affected products
HeartStart MRX Rechargeable Lithium Ion Battery
Reason
Certain MRx M3538A lithium-ion batteries may contain a defective component. Should the component fail, the battery will no longer charge or deliver power to the MRx Monitor/Defibrillator, which may prevent operation of the device if a second charged battery is not installed or if the device is not connected to AC or DC power. Additionally, the LED status indicator on the battery will not illuminate, even if the battery has a charge. Should a second, unaffected battery be present in the HeartStart MRx, the battery failure may go unnoticed.
Affected products
HeartStart MRX Rechargeable Lithium Ion Battery
Lot or serial number
Not applicable
Model or catalog number
M3538A
Companies
- Manufacturer
-
Philips Medical Systems
22100 Bothell Everett Highway
Bothell, Washington
98021
UNITED STATES