HeartStart MRx Defibrillator/Monitor (2020-07-24)
- Starting date:
- July 24, 2020
- Posting date:
- August 28, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-73759
Last updated:
2020-08-28
Affected Products
HeartStart MRx Defibrillator/Monitor
Reason
The therapy selector switch, controlled by the therapy knob, is used to turn the HeartStart MRx on in the desired mode of operation and selects the energy setting. The therapy switch may fail, resulting in the device exhibiting the following behaviors:
- The device may not perform the selected function.
- The therapy knob may not change to the energy setting selected.
- The device may deliver a shock with an energy level different from the setting selected by the user.
Affected products
HeartStart MRx Defibrillator/Monitor
Lot or serial number
Not applicable.
Model or catalog number
M3536A
Companies
- Manufacturer
-
Philips Medical Systems
22100 Bothell Everett Highway
Bothell
98021
Washington
UNITED STATES