Heartstart MRx Defibrillator/Monitor (2014-11-17)
- Starting date:
- November 17, 2014
- Posting date:
- December 4, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-42719
Recalled Products
Heartstart MRx Defibrillator/Monitor
Reason
Philips has become aware that the Heartstart MRx defibrillator/monitor may contain incorrect internal software settings, causing the following two (2) issues: 1. The device will perform the weekly automated tests hourly, which could cause the therapy capacitors to degrade sooner than intended. 2. While connected to AC or DC power and with no battery installed or the battery installed has a charge level of less than 10%, the ready for use (RFU) indicator will not provide the expected low battery indication (flashing red x with audible chirp). Instead, the RFU will show a flashing black hourglass, indicating that sufficient battery power is available for device operation. Note: once the device is disconnected from AC or DC power, the RFU indicator provides the appropriate low battery indications. In addition, all other battery charge indicators continue to operate normally, including the onscreen battery fuel gauges, low battery messages, and low battery alarms. The led charge level indicators on the batteries themselves also operate normally.
Affected products
Heartstart MRx Defibrillator/Monitor
Lot or serial number
All lots.
Model or catalog number
M3535A
M3536A
Companies
- Manufacturer
-
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES