Health product recall

Harmony™ Delivery Catheter System

Brand(s)
Last updated

Summary

Product
Harmony™ Delivery Catheter System
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Harmony™ Delivery Catheter System

More than 10 numbers, contact manufacturer.

HARMONY-DCS

Issue

Medtronic had received two (2) complaints about Harmony™ DCS distal tip detachment during the implant procedure. A root cause investigation has identified that, due to a manufacturing issue, Harmony™ DCS from specific manufacturing lots may have an increased potential for distal tip detachment under certain procedural or anatomical conditions.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies

Medtronic Inc.

710 Medtronic Parkway N.E., Minneapolis, Minnesota, United States, 55432

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82164

Get notified

Receive emails about new and updated recall and safety alerts.