Health product recall

Habib™ EndoHPB Catheter

Last updated

Summary

Product
Habib™ EndoHPB Catheter
Issue
Medical devices - Device compatibility
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Habib™ EndoHPB Catheter

All lots.

M00500070

Issue

During investigational testing of the Habib™ Catheter with the ERBE VIO 3 generator (compatible 3rd party device), an issue with the generator setting in the Habib™ IFU (50738103) was identified. From the results of this testing, it was determined that the VIO 3 setting produces results that do not meet the dimensional requirements identified for the ablation heating zone in the Habib™ design matrix (92252607).

Recall start date: March 2, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Obstetrics and gynaecology
Companies

Boston Scientific Corporation

300 Boston Scientific Way, Marlborough, Massachusetts, United States, 01752

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-73336

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe