Health product recall

GentleCool Pro Cryogen Canister (2019-09-09)

Starting date:
September 9, 2019
Posting date:
October 11, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71227



Last updated: 2019-12-11

Affected Products

GentleCool Pro Cryogen Canister

Reason

The issue observed in the field and documented in Candela's associated complaints details that the neck bushing (or top) of the canister becoming disconnected from the body of the canister, causing rapid depressurization. This event can injure the patient/operator by a) uncontrolled cryogen gas spray and b) propelling the gas canister away from the system in an uncontrolled manner with significant velocity. Please refer to the attached health hazard evaluation for a complete description.

Affected products

GentleCool Pro Cryogen Canister

Lot or serial number
  • More than 10 numbers, please contact the manufacturer.
Model or catalog number

Not applicable.

Companies
Manufacturer

Candela Corporation

530 Boston Post road

Wayland

01778

Massachusetts

UNITED STATES