GentleCool Pro Cryogen Canister (2019-09-09)
- Starting date:
- September 9, 2019
- Posting date:
- October 11, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-71227
Last updated: 2019-12-11
Affected Products
GentleCool Pro Cryogen Canister
Reason
The issue observed in the field and documented in Candela's associated complaints details that the neck bushing (or top) of the canister becoming disconnected from the body of the canister, causing rapid depressurization. This event can injure the patient/operator by a) uncontrolled cryogen gas spray and b) propelling the gas canister away from the system in an uncontrolled manner with significant velocity. Please refer to the attached health hazard evaluation for a complete description.
Affected products
GentleCool Pro Cryogen Canister
Lot or serial number
- More than 10 numbers, please contact the manufacturer.
Model or catalog number
Not applicable.
Companies
- Manufacturer
-
Candela Corporation
530 Boston Post road
Wayland
01778
Massachusetts
UNITED STATES