This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Gemstar Docking Station (2012-12-08)

Starting date:
December 8, 2012
Posting date:
March 18, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-38249

Recalled Products

Gemstar Docking Station 

Reason

The Gemstar stations do not meet the Hospira product specification for output voltage. Under certain conditions, the potential exists for the pump to fail to power up while connected to the docking station. Moreover under a specific configuration, the pump will display error code 11/003 indicating excessive input voltage from the external sources. This code will stop infusion if the error occurs while an infusion is in progress.

Affected products

Gemstar Docking Station 

Lot or serial number

All serial numbers

Model or catalog number

13075

13075-XX-YY

Companies
Manufacturer

Hospira Inc.

275 North Field Drive

Lake Forest

60045

Illinois

UNITED STATES