GEMINI TF PET/CT IMAGING SYSTEM (2015-07-01)
- Starting date:
- July 1, 2015
- Posting date:
- July 21, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-54320
Affected Products
- GEMINI TF 64 PET/CT SYSTEM - MAIN UNIT
- GEMINI TF BIG BORE PET/CT IMAGING SYSTEM
Reason
Issue #1: When a single contour is saved in edit mode, TumorLoc does not save the contour in the correct location. The contour is saved in the location of the same slice where the last contour was drawn for the last region of interest (ROI) in the (ROI) list. A misplaced contour then results in a misplaced Isocenter or incorrect field size. This occurs when using the save button, exit button or send to laser button.
Issue #2: When the region of interest (ROI) is displayed on the Beam's Eye Viewer (BEV) in outline mode, and the (ROI) color is changed, the color is changed in the Multi-Planar Projection (MPR) viewers but not the (BEV) viewers.
Issue #3: TumorLoc created datasets are flipped and incorrectly labeled when non-gated data is loaded along with pulmonary gated data and is used to create intensity projection datasets (MIP, MINIP, or AVGIP). If the clinician obtains a clinical intensity projection dataset in a certain manner, a dataset can be generated that is flipped left to right. The resulting dataset is flipped and incorrectly labeled left-to-right when displayed in a viewer.
Issue #4: Duplicate region of interest (ROI) or Isocenter names can be created in TumorLoc. This duplication is only possible when using the copy feature or manually naming a (ROI) with the suffix "copy". This duplication can be created when creating a new Isocenter or (ROI), then creating another new Isocenter or (ROI) and naming the second Isocenter or (ROI) "first name copy" (either manually or by clicking copy on the first Isocenter or (ROI), where "first name" is the name of the first Isocenter or (ROI). If the user then selects the first Isocenter or (ROI) and clicks copy, the resulting copy has the same name as the second Isocenter/ROI: "first name copy".
Affected products
A. GEMINI TF 64 PET/CT SYSTEM - MAIN UNIT
Lot or serial number
7103
Model or catalog number
4535 679 94741
Companies
- Manufacturer
-
Philips Medical Systems (Cleveland) Inc.
595 MINER ROAD
CLEVELAND
Ohio
UNITED STATES
B. GEMINI TF BIG BORE PET/CT IMAGING SYSTEM
Lot or serial number
9224
Model or catalog number
4598 001 18561
Companies
- Manufacturer
-
Philips Medical Systems (Cleveland) Inc.
595 MINER ROAD
CLEVELAND
Ohio
UNITED STATES