This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Gas/Water Feeding Valves (2014-07-14)

Starting date:
July 14, 2014
Posting date:
September 24, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-41485

Recalled products

  1. Gas/Water Feeding  Valves 

Reason

During a colonoscopy, patient experienced a perforation of the colon. Surgery was performed. After the colonoscopy the devices were flushed as per normal routine. During the after-use process it was observed that the CO2 flow through the colono scope couldn't be stopped by the CO2 valve of-B194. Patient subsequently passed away.

Affected products

A. Gas/Water Feeding  Valves

Lot or serial number

All lots

Model or catalog number
  • OF-B130
  • OF-B194
Companies
Manufacturer

Hoya Corporation

2-7-5 Naka-Ochiai, Shinjuku-ku,

Tokyo, Tokyo-to

161-8525