This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

GASTROCCULT 40T (2016-05-02)

Starting date:
May 2, 2016
Posting date:
May 16, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-58410

Affected Products

GASTROCCULT 40T

Reason

Beckman Coulter has confirmed an issue where the pH test area for the Gastroccult kit may display an unexpected color for a sample at a pH 7 reading when compared to the associated section of the color comparator bar. A sample with a pH 7 may exhibit a brownish-yellow color reading instead of the expected olive green color. The issue was first reported by a US account in March 2015 but the issue was escalated due to trending observed in January 2016.

Affected products

GASTROCCULT 40T

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 66040
Companies
Manufacturer

Beckman Coulter, Inc.

250 S. Kraemer Blvd.

Brea

92821

California

UNITED STATES