This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Gap Endo-Exo Medullary Nail Driver

Starting date:
August 10, 2017
Posting date:
September 8, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-64448

Reason

It appears that breakage of the instrument occurs when bending, impaction or excessive torsinal forces are applied on the instrument. The recalled lot numbers were previously reworked and they appear to be more susceptible to breakage considering that residual stresses may have been generated during the rework.

Affected products

Gap Endo-Exo Medullary Nail Driver

Lot or serial number

  1. 110922-052
  2. 130102-010

Model or catalog number

GAP-NDR100

Companies

Manufacturer

Pega Medical Inc.

111 autoroute Chomedey

Laval

H7W 5J8

Quebec

CANADA