Fusion Lithotripsy Extraction Basket
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Fusion Lithotripsy Extraction Basket |
Not applicable. |
FS-LXB-3X6 |
Issue
An investigation that confirmed nonconforming baskets were manufactured/distributed. This can appear in field as inability to advance or retract basket or drive wire breaking or separating from handle (with or without mechanical lithotripsy). Investigation found that bit on electric torque wrench was stripped preventing wrench from applying sufficient torque to cannula to secure drive wire. Faulty torque wrench was removed from production line and operators were retrained on torquing process.
Recall start date: July 10, 2023
Additional information
Details
| Wilson-Cook Medical, Inc. |
| 4900 Bethania Station Rd., Winston Salem, North Carolina, United States, 27105 |
Get notified