This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Fresenius Dialysate Systems (2016-12-21)

Starting date:
December 21, 2016
Posting date:
January 18, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-61828

Affected Products

  1. Fresenius 2008K Dialysate Delivery System
  2. Fresenius 2008K2 Dialysate Delivery System
  3. Fresenius 2008K2 Fifth Module Dialysate Delivery System
  4. K@Home
  5. Fresenius 2008T Hemodialysis System
  6. Fresenius 2008K2@Home
  7. Fresenius 2008T Hemodialysis System CDX
  8. 2008T Hemodialysis System CDX - Refurbished
  9. 2008K@Home - Refurbished
  10. Fresenius 2008K2 Dialysate Delivery System

Reason

The purpose of this recall is to inform users of the 2008 series hemodialysis machines of a software issue pertaining to the ultrafiltration (UF) rate during treatment. It was discovered that when the UF rate, goal or time is adjusted using the "up" and "down" arrow keys of the hemodialysis machine and the change is cancelled using the "escape" key without pressing the "confirm" key first, the "cancelled" UF rate is actually being executed rather than the rate displayed on the machine.

Affected products

A. Fresenius 2008K Dialysate Delivery System

Lot or serial number

All serial numbers

Model or catalog number
  • 190300
  • 190301
  • 190302
  • 190303
  • 190304
  • 190305
  • 190306
  • 190336
  • 190371
Companies
Manufacturer

Fresenius USA Inc.

4040 Nelson Avenue

Concord

94520

California

UNITED STATES



B. Fresenius 2008K2 Dialysate Delivery System

Lot or serial number

All serial numbers

Model or catalog number
  • 190373
  • 190452
Companies
Manufacturer

Fresenius USA Inc.

4040 Nelson Avenue

Concord

94520

California

UNITED STATES



C. Fresenius 2008K2 Fifth Module Dialysate Delivery System

Lot or serial number

All serial numbers

Model or catalog number
  • 190610
  • 190630
Companies
Manufacturer

Fresenius USA Inc.

4040 Nelson Avenue

Concord

94520

California

UNITED STATES



D. K@Home

Lot or serial number

All serial numbers

Model or catalog number
  • 190633
Companies
Manufacturer

Fresenius USA Inc.

4040 Nelson Avenue

Concord

94520

California

UNITED STATES



E. Fresenius 2008T Hemodialysis System

Lot or serial number

All serial numbers

Model or catalog number
  • 190395
  • 191014
Companies
Manufacturer

Fresenius USA Inc.

4040 Nelson Avenue

Concord

94520

California

UNITED STATES



F. Fresenius 2008K2@Home

Lot or serial number

All serial numbers

Model or catalog number
  • 190729
Companies
Manufacturer

Fresenius USA Inc.

4040 Nelson Avenue

Concord

94520

California

UNITED STATES



G. Fresenius 2008T Hemodialysis System CDX

Lot or serial number

All serial numbers

Model or catalog number
  • 190653
Companies
Manufacturer

Fresenius USA Inc.

4040 Nelson Avenue

Concord

94520

California

UNITED STATES



H. 2008T Hemodialysis System CDX - Refurbished

Lot or serial number

All serial numbers

Model or catalog number
  • 190713
  • 190766
  • 190908
  • R190713
  • R190766
Companies
Manufacturer

Fresenius USA Inc.

4040 Nelson Avenue

Concord

94520

California

UNITED STATES



I. 2008K@Home - Refurbished

Lot or serial number

All serial numbers

Model or catalog number
  • RTLR190573
Companies
Manufacturer

Fresenius USA Inc.

4040 Nelson Avenue

Concord

94520

California

UNITED STATES



J. Fresenius 2008K2 Dialysate Delivery System

Lot or serial number

All serial numbers

Model or catalog number
  • R190395
Companies
Manufacturer

Fresenius USA Inc.

4040 Nelson Avenue

Concord

94520

California

UNITED STATES