Freelite Human Lambda Free Kit (for use on the SPAPLUS)
Brand(s)
Summary
Immediately stop using the product and contact manufacturer. |
Affected products
Affected Products |
Lot or serial number |
Model or catalog number |
---|---|---|
FREELITE HUMAN LAMBDA FREE KIT,USE ON THE SPA PLUS |
Lot: 435251 |
LK 018.S |
Issue
Following an investigation of customer complaints, an increase in curve activity and decrease in top point separation between calibrator (CAL) points 5 and 6 was observed for IK018.S lot 472776 (lK018.L.S lot 435251 shares the same components). Cal. 5 OD values may occasionally exceed Cal.6 Values, which may lead to imprecision within this region of the curve. Therefore, TBS will initiate the correction by way of issuing an UFSN to the affected customers.
Recall date: 2021-10-14 |
What you should do
Immediately stop using the product and contact manufacturer. |
Additional information
Details
THE BINDING SITE GROUP LIMITED 8 CALTHORPE ROAD, EDGBASTON, B15 1QT
|