Issue:
Medical Devices
Audience:
Hospitals, General Public, Healthcare Professionals
Identification number:
RA-64836

Affected products

FREE TESTOSTERONE ELISA KIT

Reason

Free testosterone ELISA kits, lot 171390, may fail to produce any optical density (OD) values and therefore any results. The low OD values would lead to an inability to meet the QC certificate acceptance criteria. This would result in an invalid test and therefore, no results would be reported.