This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

FORMULA THERAPY SYSTEMS (2016-10-17)

Starting date:
October 17, 2016
Posting date:
October 28, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60786

Affected Products

Reason

This voluntary field corrective action is being taken on the formula family of hemodialysis machines following reports of three thermal events which led to fire, involving three different machines in the field. No patients or healthcare providers were injured as a result of these events. Preliminary investigations indicate that the battery charge board was the source of two of the events. Due to extensive damage on the machine, the source of the third event is difficult to conclude. Investigation into the trigger for the events is on-going. This recall involves all machines manufactured and imported since 1998. The corrective action will involve disconnection of the battery charge board to eliminate the possibility of the failure mode while a permanent solution is implemented.

Affected products

A. FORMULA 2000 PLUS DOMUS

Lot or serial number

All serial numbers imported

Model or catalog number

IBFPBD301

Companies
Manufacturer

Bellco S.R.L.

Via Camurana 1

Mirandola, Modena

41037

ITALY



B. FORMULA DOMUS

Lot or serial number

All serial numbers imported

Model or catalog number

IBFXMD301

IBFXMD310

Companies
Manufacturer

Bellco S.R.L.

Via Camurana 1

Mirandola, Modena

41037

ITALY



C. FORMULA

Lot or serial number

All serial numbers imported

Model or catalog number

IBFXMX301

IBFXMX310

IBFXMX700

Companies
Manufacturer

Bellco S.R.L.

Via Camurana 1

Mirandola, Modena

41037

ITALY



D. FORMULA 2000

Lot or serial number

All serial numbers imported

Model or catalog number

IBFXBX301

IBFXBX310

IBFXBX700

Companies
Manufacturer

Bellco S.R.L.

Via Camurana 1

Mirandola, Modena

41037

ITALY



E. FORMULA 2000 DOMUS

Lot or serial number

All serial numbers imported

Model or catalog number

IBFPBD37D

IBFXBD301

IBFXBD310

Companies
Manufacturer

Bellco S.R.L.

Via Camurana 1

Mirandola, Modena

41037

ITALY



F. FORMULAR 2000 PLUS

Lot or serial number

All serial numbers imported

Model or catalog number

IBFPBX301

IBFPBX310

IBFPBX37D

Companies
Manufacturer

Bellco S.R.L.

Via Camurana 1

Mirandola, Modena

41037

ITALY



G. FORMULA THERAPY

Lot or serial number

All serial numbers imported

Model or catalog number

IBFTBX301

Companies
Manufacturer

Bellco S.R.L.

Via Camurana 1

Mirandola, Modena

41037

ITALY



H. FORMULA THERAPY DOMUS

Lot or serial number

All serial numbers imported

Model or catalog number

IBFTBD301

Companies
Manufacturer

Bellco S.R.L.

Via Camurana 1

Mirandola, Modena

41037

ITALY