Health product recall

FLOVAC SUCTION/IRRIGATION CONTROL HANDLES (2016-06-29)

Starting date:
June 29, 2016
Posting date:
July 22, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Hospitals, Healthcare Professionals
Identification number:
RA-59470

Affected products

A. FLOVAC SUCTION/IRRIGATION CONTROL HANDLES

Reason

The core suction irrigation handpieces are sold as sterile, single use devices. ConMed has identified certain core suction irrigation product with creases in the packaging seal which sometimes result in an open channel. These channels may compromise sterility of the product. In no instance has it been reported to ConMed that a compromise in the sterile barrier has resulted in illness or injury.

Affected products

A. FLOVAC SUCTION/IRRIGATION CONTROL HANDLES

Lot or serial number

110609X - 20160304X

Model or catalog number

CD8185

CD8190

CD8200

CD8300

CD8302

CD8400

CD8450

Companies
Manufacturer

Conmed Corporation

525 French Road

Utica

13502

New York

UNITED STATES