Health product recall

FilmArray GI Panel (2019-03-07)

Starting date:
March 7, 2019
Posting date:
April 5, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-69474


Last updated:

Affected Products

FilmArray GI Panel

Reason

BioFire Diagnostics, LLC (BioFire) has identified elevated rates of false positive results for Campylobacter and Cryptosporidium while using the BioFire GI Panel. Only the BioFire GI Panel kits with expiration dates ranging from 10/23/2019 to 02/20/2020 are affected.  If using a BioFire GI Panel kit with an expiration date that is mentioned above, positive results for Campylobacter and Cryptosporidium should be confirmed by another method prior to reporting the test results.

Affected products

FilmArray GI Panel

Lot or serial number
  • 365818
  • 381618
  • 397218
  • 510618
  • 523918
Model or catalog number
  • RFIT ASY-0104
  • RFIT ASY-0116
Companies
Manufacturer

BioFire Diagnostics, LLC

515 Colorow Dr

Salt Lake City

84108

Utah

UNITED STATES