Fetalscreen II
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Fetalscreen II |
V273985 |
Z488 |
Issue
A customer complaint investigation has confirmed that the lot of Fetalscreen II stated above shows unexpected macroscopic reactivity (agglutination) when performing QC testing with the positive control cells. This will impact the reading of rosettes to validate the positive control and may impact the reading of positive tests, potentially resulting in false negative results.
Recall Start Date: July 22, 2024
Additional information
Details
Alba Bioscience Limited
James Hamilton Way, Penicuik, Scotland, Great Britain, EH26 OBF
Get notified