Health product recall

FDR Visionary Suite

Last updated

Summary

Product
FDR Visionary Suite
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare
Industry

Affected products

Affected products Lot or serial number Model or catalogue number
FDR Visionary Suite Serial number: MP95A9E9C001 MP95A9E9C001

Issue

Potential safety issue with the X-ray high voltage generator incorporated into the FDR Visionary Suite X-ray diagnostic equipment. It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator may weld, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it is possible that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Shimadzu Corporation

1, Nishinokyo-Kuwabaracho, Nakagyo-Ku, Kyoto-Fu, Japan, 604-8511

Published by
Health Canada
Audience
Healthcare
Industry
Recall class
Type II
Recall date
Identification number
RA-81686