Health product recall

EZ Glide Cannula, Straight, Curved With Suture Bump & Curved With Suture Flange (2019-10-24)

Starting date:
October 24, 2019
Posting date:
November 1, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71465

Last updated: 2019-10-29

Affected Products

A. EZ Glide Cannula, Straight
B. EZ Glide Cannula, Curved With Suture Bump
C. EZ Glide Cannula, Curved With Suture Flange

Reason

Edwards Lifesciences has initiated a recall of the EZ Glide product after receiving three (3) reports that an EZ Glide Cannula separated from its connector, causing a breach of the cardiopulmonary bypass (CPB) circuit and loss of blood. In each case, the report suggests that the separation occurred without any significant force being applied to the joint. Although the occurrence rate is 0.0034%, And all patients had successful surgical outcomes, in the interests of patient safety and transparency, Edwards is notifying customers of the events and requesting return of EZ Glide devices.

Affected products

A. EZ Glide Cannula, Straightj

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • EZS21A
  • EZS21TA
  • EZS24A
  • EZS24TA
Companies
Manufacturer

Edwards Lifesciences LLC

1 Edwards Way

Irvine

92614

California

UNITED STATES

B. EZ Glide Cannula, Curved With Suture Bump

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • EZC21A
  • EZC21TA
  • EZC24A
  • EZC24TA
Companies
Manufacturer

Edwards Lifesciences LLC

1 Edwards Way

Irvine

92614

California

UNITED STATES

C. EZ Glide Cannula, Curved With Suture Flange

Lot or serial number

More than 10 numbers, contact manufacturer. 

Model or catalog number
  • EZF21A
  • EZF21TA
  • EZF24A
  • EZF24TA
Companies
Manufacturer

Edwards Lifesciences LLC

1 Edwards Way

Irvine

92614

California

UNITED STATES