Exoplan
Brand(s)
Last updated
Summary
Product
Exoplan
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Exoplan |
(01)4260521365019(10)A03B00E** |
2.3 MATERA |
Issue
During an update of data used in a library shipped with Exoplan or made available on the download portal of Exocad, it was found that the use of a specific combination of an implant, sleeve and surgical drill can result in a hole being drilled 1 mm deeper than expected.
Recall start date: October 12, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
| Exocad Gmbh |
| Rosa-Parks-Str.2, Darmstadt, Germany, 64295 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74654
Get notified