This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

EVOTECH Endoscope Cleaner & Reprocessor (2015-05-14)

Starting date:
May 14, 2015
Posting date:
June 10, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-53737

Recalled Products

A. EVOTECH Endoscope Cleaner & Reprocessor

Reason

Advanced Sterilization Products recently determined that the Evotech endoscope cleaner and reprocessor (ECR) may not detect an improper endoscope connection on two channels, and subsequently pass the cycle without cancellation if there is a disconnected scope. This is not a removal of the Evotech ECR and customers may continue to use the device in accordance with the user's guide until a system correction is available.

Affected products

A. EVOTECH Endoscope Cleaner & Reprocessor

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

50014-101                       

50014-102

Companies
Manufacturer

Advanced Sterilization Products

A Division of Ethicon

33 Technology Drive

Irvine

92618

UNITED STATES