Evolve Triad Plate Cutter
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Evolve Triad Plate Cutter |
2486857 |
49510120 |
Issue
Stryker has identified a nonconformance relating to the Evolve Triad Plate Cutter provided in the Evolve Triad Instrument Kit (4951kit1). Specifically, the plate cutter is unable to meet sterilization requirements when it is sterilized in a double wrapped configuration according to the option 1 steam sterilization parameters detailed in Instructions for Use (IFU150824). While the surgical technique does not require use of the plate cutter, it is provided in the instrument kit so the coronoid plate can be trimmed when needed.
Recall start date: June 3, 2022
Additional information
Details
Wright Medical Technology Inc.
1023 Cherry Road, Memphis, Tennessee, United States, 38117
Get notified
Receive emails about new and updated recall and safety alerts.