This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

EVOLUTION MP Total Knee Sys - Tibial Base Keeled (2015-01-28)

Starting date:
January 28, 2015
Posting date:
March 6, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-43809

Recalled products

  1. Evolution Tibial Base

Reason

Device in question has varying wall thickness and keel offset greater than the maximum allowable tolerance due to a casting issue

Affected products

A. Evolution Tibial Base

Lot or serial number
  • 1560140
Model or catalog number
  • ETPKN8SR
Companies
Manufacturer

Microport Orthopedics Ltd.

5677 Airline Road

Arlington

38002

Texas

UNITED STATES