Health product recall

ETEST ANTIMICROBIAL SUSCEPTIBILITY TESTING - COLI (2016-07-26)

Starting date:
July 26, 2016
Posting date:
August 19, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-59854

Affected products

  1. ETEST ANTIMICROBIAL SUSCEPTIBILITY TESTING - COLI

Reason

Based on scientific publications (including EUCAST warning letter on their website) on colistin gradient and disk diffusion tests, including Etest colistin, bioMerieux was alerted about a potential underestimation of the MIC values for entero bacteriaceae and acinetobacter. bioMerieux initiated a complaint investigation to confirm product performance and determine root cause. While the internal and external investigations are still ongoing.

Affected products

A. ETEST ANTIMICROBIAL SUSCEPTIBILITY TESTING - COLI

Lot or serial number

More than 10 numbers, contact manufacturer.  

Model or catalog number

537300
537308

Companies
Manufacturer
Biomerieux Sa
Chemin De L'orme
Marcy-l'etoile
69280
FRANCE