This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

ETEST ANTIMICROBIAL SUSCEPTIBILITY TESTING - COLI (2016-07-26)

Starting date:
July 26, 2016
Posting date:
August 19, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-59854

Affected products

  1. ETEST ANTIMICROBIAL SUSCEPTIBILITY TESTING - COLI

Reason

Based on scientific publications (including EUCAST warning letter on their website) on colistin gradient and disk diffusion tests, including Etest colistin, bioMerieux was alerted about a potential underestimation of the MIC values for entero bacteriaceae and acinetobacter. bioMerieux initiated a complaint investigation to confirm product performance and determine root cause. While the internal and external investigations are still ongoing.

Affected products

A. ETEST ANTIMICROBIAL SUSCEPTIBILITY TESTING - COLI

Lot or serial number

More than 10 numbers, contact manufacturer.  

Model or catalog number

537300

537308

Companies
Manufacturer

Biomerieux Sa

Chemin De L'orme

Marcy-l'etoile

69280

FRANCE