This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

ETEST ANTIMICROBIAL SUSCEPTIBILITY TESTING (2015-11-26)

Starting date:
November 26, 2015
Posting date:
January 14, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-56596

Affected Products

A. ETEST ANTIMICROBIAL SUSCEPTIBILITY TESTING - PIPERACILLIN/TAZOBACTAM

B. ETEST ANTIMICROBIAL SUSCEPTIBILITY TESTING - PIP/TAZO/CON-4

Reason

On October 27th, 2015, a customer reported a low Mic on Etest Piperacillin/Tazobactam (4 ug/ml) PTC-256 single pack (lot 1004271120). The customer obtained erroneous results (false susceptible) on Etest with broth-micro dilution reference method.

Affected products

A. ETEST ANTIMICROBIAL SUSCEPTIBILITY TESTING - PIPERACILLIN/TAZOBACTAM

Lot or serial number
  • 1002092270
  • 1002276210
  • 1002375410
  • 1002482640
  • 1002995820
  • 1003105180
  • 1003416920
  • 1003699550
  • 1003980200
  • 1004180840
  • 1004276460
  • 1004398470
Model or catalog number

521418

Companies
Manufacturer

Biomerieux Inc.

595 Anglum Road

Hazelwood

63042

Missouri

UNITED STATES



B. ETEST ANTIMICROBIAL SUSCEPTIBILITY TESTING - PIP/TAZO/CON-4

Lot or serial number
  • 1001936360
  • 1002488490
  • 1003069230
  • 1003276060
  • 1003415090
  • 1003697470
  • 1003977300
  • 1004178060
  • 1004271120
  • 1004331500
Model or catalog number

412434

Companies
Manufacturer

Biomerieux Inc.

595 Anglum Road

Hazelwood

63042

Missouri

UNITED STATES