Health product recall

epoc® NXS Host – 5553CA

Brand(s)
Last updated

Summary

Product
epoc® NXS Host – 5553CA
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

epoc® NXS Host – 5553CA

More than 10 numbers, contact manufacturer.

11413506

Issue

Siemens Healthcare Diagnostics Inc. has confirmed a potential issue that could occur under specific conditions when the (automatic) delete blood tests feature is enabled. When the results are run on the epoc® NXS host and delete blood tests feature is enabled, the results that are generated at the time of testing are correct; however, when this issue occurs, results that are printed, saved and/or transferred to a data manager, may include unselected analytes that are from a previous patient or QC test.

Recall start date: October 24, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Epocal Inc.

2060 Walkley Road, Ottawa, Ontario, Canada, K1G 3P5

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74607

Get notified

Receive notifications for new and updated recalls and alerts by category.