Health product recall

epoc® Blood Analysis BGEM Test Cards

Brand(s)
Last updated

Summary

Product
epoc® Blood Analysis BGEM Test Cards
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare
Industry

Affected products

Affected products

Lot or serial number

Model or catalog number

epoc® Blood Analysis BGEM Test Cards

More than 10 numbers, contact manufacturer.

10736515

Issue

Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in some epoc® test card lots. All other analytes are performing as intended. The observed average bias for sodium was -4.4 mmol/l. The maximum bias observed was -14 mmol/l, which occurred at a higher sodium concentration around 150 mmol/l. No positive bias has been observed. Quality control (qc) fluid may or may not detect the issue depending on the bias magnitude and the frequency of qc testing.

Recall start date: July 9, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies

Epocal Inc.

2060 Walkley Road, Ottawa, Ontario, Canada, K1G 3P5

Published by
Health Canada
Audience
Healthcare
Industry
Recall class
Type II
Identification number
RA-77740