This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Engstrom Carestation (2015-01-29)

Starting date:
January 29, 2015
Posting date:
February 6, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-43635

Recalled products

  1. Engstrom Carestation

Reason

Specific combinations of Engstrom ventilator casters and bases have been identified that pose an increased likelihood of failures of the caster mounting hardware. A failure of the caster mounting hardware may result in the ventilator falling/tipping (overbalancing). A fall/tip of the Engstrom ventilator can result in serious injury to patients or staff.

Affected products

A. Engstrom Carestation

Lot or serial number
  • CBCP01415
  • CBCP01452
  • CBCP01475
Model or catalog number
  • 1505-9000-000
Companies
Manufacturer

Datex-Ohmeda

3030 Ohmeda Drive, P.O. Box 7550,

Madison

53707-7550

Wisconsin

UNITED STATES