Embryology Heated Plate within RI Witness (2019-09-10)
- Starting date:
- September 10, 2019
- Posting date:
- October 25, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-71353
Last updated: 2019-10-25
Affected Products
Embryology Heated Plate within RI Witness
Reason
CooperSurgical became aware from complaint investigations that there is a risk that the touchpad may not work properly after cleaning of the Embryology Heated Plate within the RI Witness. The root cause has been identified to be liquid ingress into the keypad which has the potential to produce an electrical path across the tracks on the keypads' membrane tail. This failure mode does not affect the temperature controller, which continues to maintain proper temperature.
Affected products
Embryology Heated Plate within RI Witness
Lot or serial number
- 603
- 604
- 611
- 614
- 633
- 635
Model or catalog number
- 6-70-807
- 6-70-807-A
- 6-70-807-B
Companies
- Manufacturer
-
Coopersurgical Inc.
95 Corporate Drive
Trumbull
06611
Connecticut
UNITED STATES