Health product recall

Embryology Heated Plate within RI Witness (2019-09-10)

Starting date:
September 10, 2019
Posting date:
October 25, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71353

Last updated: 2019-10-25

Affected Products

Embryology Heated Plate within RI Witness

Reason

CooperSurgical became aware from complaint investigations that there is a risk that the touchpad may not work properly after cleaning of the Embryology Heated Plate within the RI Witness. The root cause has been identified to be liquid ingress into the keypad which has the potential to produce an electrical path across the tracks on the keypads' membrane tail. This failure mode does not affect the temperature controller, which continues to maintain proper temperature.

Affected products

Embryology Heated Plate within RI Witness

Lot or serial number
  • 603
  • 604
  • 611
  • 614
  • 633
  • 635
Model or catalog number
  • 6-70-807
  • 6-70-807-A
  • 6-70-807-B
Companies
Manufacturer

Coopersurgical Inc.

95 Corporate Drive

Trumbull

06611

Connecticut

UNITED STATES