Health product recall

EMBLEM S-ICD SUBCUTANEOUS ELECTRODE (2021-09-28)



Starting date:
September 28, 2021
Posting date:
October 15, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-76647


Last updated:

Affected Products

EMBLEM S-ICD SUBCUTANEOUS ELECTRODE

Reason

In December 2020, Boston Scientific notified users of the Emblem Electrode (Model 3501) about the potential for body fractures distal to the proximal sense ring. The electrode demonstrated a low malfunction rate and has continued to be available while a design enhancement was pursued. Facilities are requested to return the affected/original electrodes as the enhanced version is now available.

Affected products

EMBLEM S-ICD SUBCUTANEOUS ELECTRODE

Lot or serial number

643501-700

643501-700

Model or catalog number

3501

Companies
Manufacturer

Boston Scientific Corporation

4100 Hamline Avenue North

St. Paul

55112-5798

Minnesota

UNITED STATES