This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

ELLEX INTEGRA PRO LP5532 (2016-03-16)

Starting date:
March 6, 2016
Posting date:
April 4, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-57758

Affected Products

ELLEX INTEGRA PRO LP5532

Reason

Under certain conditions there is a potential for the glass of the safety filter optics to crack on their lower edge due to pressure applied on them by their retaining plate which in turn could damage the laser safety coating on the optics. If this occurs, there is a remote possibility the user could be exposed to any diffused reflected laser light during laser treatment.

Affected products

ELLEX INTEGRA PRO LP5532

Lot or serial number

SN # 5A2058

SN # 5A2067

Model or catalog number

LP5532

Companies
Manufacturer

ELLEX MEDICAL PTY LTD

82 GILBERT STREET

ADELAIDE

5000

AUSTRALIA