Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-43483

Recalled products

  1. Elekta Precise Treatment System

Reason

Elekta has identified a batch of securing bolts that fix the gantry counter frame weight stack may fail; the potentially faulty bolts are marked with "OF" on the top. The gantry counter frame weight stack is secured with 6 bolts. The original customer complaint reported that two of the bolts on the machine holding the counter frame sheared within a week of installation.