Health product recall

Elecsys® Anti-TSHR

Last updated

Summary

Product
Elecsys® Anti-TSHR
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Elecsys® Anti-TSHR

840177

08496609190

Elecsys® Anti-TSHR

840183

8496633190

Elecsys® Anti-TSHR

874011

8496633190

Issue

Internal reviews have confirmed two related problems with the Elecsys® Anti-TSHR assay. The test results for the same patient sample vary by instrument platforms (Cobas e 801/e 402 and Cobas e 411) or results vary by calibrator lot (only affecting Cobas e 411 platform). Roche will replace all of the impacted reagents still remaining in your inventory, free of charge. All newly manufactured lots are not impacted.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies

Roche Diagnostics GmbH

Sandhofer Strasse 116, Mannheim, Germany, 68305

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-81394

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe