This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Elecsys HIV Combi PT (2014-01-29)

Starting date:
January 29, 2014
Posting date:
April 14, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-38987

Recalled Products

Elecsys HIV Combi PT

Reason

Manufacturer has identified through customers complaints a possibility that the Elecsys HIV Combi PT may show a calibration signal shift in the control, PC HIV level 1, and may read out of range (greater than 3SD).

Affected products

Elecsys HIV Combi PT

Lot or serial number

171983

172659

173326

175019

Model or catalog number

05390095190

Companies
Manufacturer

Roche Diagnostics GmbH

Sandhofer Strasse 116

Mannheim

68305

GERMANY