This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Elecsys FT4
- Starting date:
- October 18, 2012
- Posting date:
- November 26, 2012
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices, Product Safety
- Audience:
- General Public (GP), Healthcare Professionals, Hospitals
- Identification number:
- RA-15901
Recalled Products
- Elecsys FT4
- Manufacturer
-
Roche Diagnostics GMBH
Reason
During recent internal on-board stability studies, decreasing calibrator signals were observed for Elecsys FT4 reagent kits when stored more than 3 weeks on-board the Elecsys systems (Cobas E 411 and Cobas E 601/602)
Affected products
A. Elecsys FT4
Lot or serial number
16519303, 16519304, 16649801, 16649803
16763501, 16763502, 16763503, 16845401, 16845403
Model or catalog number
11731297122
Companies