Health product recall

Elecsys Anti-TSHR (2018-08-21)

Starting date:
August 21, 2018
Posting date:
January 11, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68734

Affected products

Elecsys Anti-TSHR

Reason

Following internal investigation of customer complaints, the manufacturer confirmed a limitation in On-Board Stability (OBS) of the product Elecsys Anti-TSHR (#07026951190) lot #28356401 used on cobas e 801 module. The OBS could be maintained for 8 weeks, in comparison to the 16 weeks OBS claim as described in the Method Sheet of the assay.

Affected products

Elecsys Anti-TSHR

Lot or serial number

28356401

Model or catalog number

07026951190

Companies
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY