Health product recall

Eclipse Treatment Planning System

Starting date:
August 17, 2012
Posting date:
September 17, 2012
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public (GP), Healthcare Professionals, Hospitals
Identification number:
RA-15530

Recalled Products:

Eclipse Treatment Planning System

Reason

Eclipse treatment planning system where an outdated assigned Hounsfield Unit [Hu] value can be used for dose calculation when the clock on the Eclipse Client workstation is not synchronized with the database / system server. When the Eclipse client workstation clock is running behind the clock on the database / system server, it is possible that a change to the assigned HU value for a structure does not invalidate the density image stored in the cache, and for the subsequent dose and Monitor Unit [MU] calculation to be based on the outdated HU assignment. The resultant MUs may be higher or lower than expected, dependent on the specific HU assignments. Treatment of the patient using these values can therefore lead to under or over dose.

Affected products

Eclipse Treatment Planning System

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

100042024

Companies
Manufacturer
Varian Medical Systems Inc.