This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Eclipse Treatment Planning System
- Starting date:
- August 17, 2012
- Posting date:
- September 17, 2012
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public (GP), Healthcare Professionals, Hospitals
- Identification number:
- RA-15530
Recalled Products:
Eclipse Treatment Planning System
Reason
Eclipse treatment planning system where an outdated assigned Hounsfield Unit [Hu] value can be used for dose calculation when the clock on the Eclipse Client workstation is not synchronized with the database / system server. When the Eclipse client workstation clock is running behind the clock on the database / system server, it is possible that a change to the assigned HU value for a structure does not invalidate the density image stored in the cache, and for the subsequent dose and Monitor Unit [MU] calculation to be based on the outdated HU assignment. The resultant MUs may be higher or lower than expected, dependent on the specific HU assignments. Treatment of the patient using these values can therefore lead to under or over dose.
Affected products
Eclipse Treatment Planning System
Lot or serial number
More than 10 numbers, contact manufacturer
Model or catalog number
100042024
Companies
- Manufacturer
-
Varian Medical Systems Inc.