Health product recall

Duo Flow Double Lumen and Soft-Line Catheter Sets, Raulerson Pre-Curved Duo-Flow IJ Catgeter With Soft Tip (2018-07-13)

Starting date:
July 13, 2018
Posting date:
July 27, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67318

Affected products

A. Duo Flow Double Lumen Catheter Sets
B. Raulerson Pre-Curved Duo-Flow IJ Catgeter With Soft Tip
C. Soft-Line Catheter Sets

Reason

A recall has been initiated for various products in the short term hemodialysis line due to the potential for the suture wings to crack or break. No patient injury has been reported. Implanted catheters with intact suture wings do not need to be removed. However, an alternative method of catheter securement should be initiated for all implanted catheters. Potential patient injury includes catheter dislocation, bleeding and air emboli.

Affected products

A. Duo Flow Double Lumen Catheter Sets

Lot or serial number

MCBF000
MCBG700
MCDP270
MCFF150
MCFF580
MCFF780
MKAH230
MKAH580

Model or catalog number

XTP114MT
XTP94MT
XTP96MT

Companies
Manufacturer

Medical Components Inc,

Also trading as Medcomp

1499 Delp Drive

Harleysville

19438

Pennsylvania

UNITED STATES

B. Raulerson Pre-Curved Duo-Flow IJ Catgeter With Soft Tip

Lot or serial number

MCDT880

Model or catalog number

XTP94IJS

Companies
Manufacturer

Medical Components Inc,

Also trading as Medcomp

1499 Delp Drive

Harleysville

19438

Pennsylvania

UNITED STATES

C. Soft-Line Catheter Sets

Lot or serial number

MCBM870
MHTH950
MJKN210
MJLP090
MKAM870
MLBA570
MLBK460
MSHN190

Model or catalog number

T116M
T94M
T96M

Companies
Manufacturer

Medical Components Inc,

Also trading as Medcomp

1499 Delp Drive

Harleysville

19438

Pennsylvania

UNITED STATES