DSX- Revelation software 6.08-6.28 (2018-08-03)
- Starting date:
- August 3, 2018
- Posting date:
- September 17, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-67764
Affected products
DSX- Revelation software 6.08-6.28
Reason
The DSX analyzer utilizes operating software Revelatlon. This software use set up as default with an algorithm for dealing with over/ under conversion. If an end user turns this off and they use the over limit in their Quality Control Equation the software will place 0.00 in the equation leading to an incorrect QC interpretation and potential incorrect value assigned to the result.
Affected products
DSX- Revelation software 6.08-6.28
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
65078-XXX
Companies
- Manufacturer
-
Dynex Technologies Inc.
14340 Sullyfield Cir
Chantilly
20151-1621
Virginia
UNITED STATES