This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

DRX-Revolution Mobile X-Ray System

Starting date:
May 11, 2017
Posting date:
June 7, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63532

Reason

Carestream Health has identified an issue with the DRX-Revolution Mobile X-Ray system that could result in separation of the tube and yoke assembly. Carestream found a single device where four screws in the assembly were not applied in the manufacturing process. To date, Carestream has found only a single system where this failure has occurred and considers it unlikely that there are other DRX-Revolution Mobile X-ray systems where this step was missed. There was no injury associated with this failure.

Affected products

DRX-Revolution Mobile X-Ray System

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

1019397

Companies

Manufacturer

Carestream Health Inc.

150 Verona Street

Rochester

14608

New York

UNITED STATES